• Overall Height:
      11.67“ (296.4mm)
    • Overall Width:
      Maxmum: 6.5“ (165.1mm)
      Minimum: 3.2“ (81.3mm)
    • Overall Depth:
      6.05“ (153.7mm)
    • Weight:
      5.25lbs (2.38kg)
    • Lifetime:
      > 5.7 Million cycles [1 cycle = 1 full dispense and aspirate move]
    • Operating Temperature:
      2°C to 50°C (36°F to 122°F)
    • Operating Humidity:
      <80% Relative Humidity at 5°C to
      30°C and <50% at 31°C to 50°C
    • Storage Temperature:
      -33°C to 71°C (-27°F to 160°F)
    • Compliance:
      REACH & RoHS Compliant
      UL & CE Compliant
      CSA Certified
    • Mounting Holes:
      Bottom
    • Resolutions:
      24K, and 48K Increments
    • Speed:
      0.0625 - 10,000 steps/sec
    • Syringe Size (60mm stroke):
      250μL to 5mL (supplied separately)
    • Valve Type:
      FAS MS-E 3/2 Solenoid Valve per channel
    • Precision:
      0.50% CV at 1/10th dispense
      0.05% CV at full stoke dispense
    • Accuracy:
      ±0.3% at 1/10th dispense
      ±0.2% at full stroke dispense
    • Communications Options:
      RS-232, RS-485, CANbus
    • Power Requirements:
      24V (with maximum power voltage ripple of 720mv peak-to-peak).
    • Power Consumption:
      Unenergized: 3.5W
      Energized: 62W (peak)